Bio-Rad Rat Diabetes Assays User Manual

Page 27

Advertising
background image

25

# of Wells

96

20x Detection

Antibodies, µl

150

Detection

Antibody

Diluent, µl

2,850

Total

Volume, µl

3,000

# of Wells

96

20x Detection

Antibodies, µl

Singleplex #1

150

20x Detection

Antibodies, µl

Singleplex #2

150

Detection

Antibody

Diluent, µl

2,700

Total

Volume, µl

3,000

# of Wells

96

20x Detection

Antibodies, µl

Diabetes

150

10x Detection

Antibodies, µl

Cytokines

300

Detection

Antibody

Diluent, µl

2,550

Total

Volume, µl

3,000

For reference, Table 15 summarizes volumes required for preparing
1x detection antibodies from a single 20x stock.

Table 16 summarizes volumes required for preparing 1x detection

antibodies from a mix of two 20x stocks; volumes required for
preparing detection antibodies of one diabetes (20x) assay and one
cytokine (10x) assay (for example, human insulin and human IL-6)
are listed in Table 17.

Preparing 1x detection antibodies from 20x stock (includes 25% excess volume).

Table 15. Premixed panel or one singleplex assay.

Table 16. Mixing two singleplex assays or a premixed panel + singleplex assay.

Table 17. Preparing 1x detection antibodies from two stocks at different concentrations.
Mixing human insulin (20x) with human IL-6 (10x) is one example.*

* Due to differences in dilution factors, it is not possible to multiplex adiponectin, adipsin

(human), VCAM-1, or ICAM-1 with other diabetes or cytokine assays.

5. After incubating the beads, samples, standards, and blank, slowly

remove and discard the sealing tape.

Advertising
This manual is related to the following products: