Bio-Rad Rat Diabetes Assays User Manual

Page 17

Advertising
background image

15

3. For high PMT/narrow range cytokine standard curve, add 12.8 μl

of the reconstituted cytokine standard and 128 μl of the
reconstituted diabetes standard to the S1 tube for a total volume
of 200 μl (Figure 4). Vortex at medium speed for 5 sec.

4. Add 150 μl of standard diluent to the remaining tubes, as shown
in Figure 4.

Fig. 4. S1 mixture and fourfold dilution series of diabetes and cytokine standards for
detection at high PMT. Produces a narrow range cytokine standard curve.

5. For low PMT/broad range cytokine standard curve, add 64 µl of each

standard to the S1 tube, total volume 200 µl (Figure 5). Vortex for 5 sec.

6. Use a new pipet tip to transfer 50 μl from the S1 tube to the S2 tube

containing standard diluent.

7. Continue with 1:4 (fourfold) serial dilutions from tube S2 to S8 as

shown in Figure 5.

Fig. 5. S1 mixture and fourfold dilution series of diabetes and cytokine standards for
detection at low PMT. Produces a broad range cytokine standard curve.

150 150 150 150 150 150 150 150

S2 S3 S4 S5 S6 S7 S8

Blank

Diluent, µl

50 50 50 50 50 50 50

Transfer

Volume, µl

250 µl

Reconstituted

Cytokine

Standard

72 µl

Standard Diluent

“S1” Tube, 200 µl Total

64 µl

64 µl

250 µl

Reconstituted

Diabetes

Standard

150 150 150 150 150 150 150 150

S2 S3 S4 S5 S6 S7 S8

Blank

Diluent, µl

50 50 50 50 50 50 50

Transfer

Volume, µl

500 µl

Reconstituted

Cytokine

Standard

59.2 µl

Standard Diluent

“S1” Tube, 200 µl Total

12.8 µl

128 µl

500 µl

Reconstituted

Diabetes

Standard