Prepare and add detection antibodies – Bio-Rad Rat Diabetes Assays User Manual

Page 26

Advertising
background image

24

Prepare and Add Detection Antibodies

Instructions are provided for diluting the detection antibodies to a 1x
concentration. When mixing diabetes and cytokine assays, keep in mind
the stock concentrations of detection antibodies as shown below.

Table 14. Stock concentration of detection antibodies.

1. While the samples are incubating, use Tables 15–17 to calculate the

volume of detection antibodies and detection antibody diluent needed.
Detection antibodies should be prepared 10 min before use.

2. Add the required volume of Bio-Plex detection antibody diluent to a

15 ml polypropylene tube.

3. Vortex the stock detection antibodies for 15–20 sec at medium

speed, then perform a 30 sec spin to collect the entire volume at the
bottom of the tube.

4. Dilute detection antibodies to 1x by pipetting the required volume into

the 15 ml tube.

Each well of the assay requires either 1.25 μl (20x stock) or 2.5 µl

(10x stock) adjusted to a final volume of 25 μl in detection
antibody diluent.

Stock Concentration of

Assay

Detection Antibodies

Human/NHP, mouse, and rat diabetes

20x

Human and mouse cytokines (groups I, II)

10x

Rat cytokines (group I)

20x

Mouse cytokines (group III)

20X

Advertising