Bio-Rad Bio-Scale™ Mini CHT™ 40 µm Cartridges User Manual

Page 11

Advertising
background image

Table 3. Products for Buffer Exchange

Sample Recommended
Volume

Product Use

Catalog

#

50–100 µl

Bio-Spin 6 column Desalting proteins

732-6000

≥6 kD

50–100 µl Bio-Spin 30 column Desalting proteins

732-6004

≥30 kD

100 µl–3 ml Bio-Scale Mini P6

Desalting proteins

732-4502

cartridge

≥6 kD

Up to 3 ml Econo-Pac 10DG

Desalting proteins

732-2010

desalting columns

≥6 kD

Unlimited

Bio-Gel P-6DG gel Desalting proteins

150-0738

≥6 kD

4.2 General Purification Protocol

Hydroxyapatite chromatography is usually performed
using increasing linear salt gradients to elute the
sample components. For best results, and increased
cartridge life, samples and buffers should be degassed
and filtered through a 0.45 µm filter. In order to
promote the stability of CHT, buffers are recommended
to include low concentrations of either phosphate or

15

5. Equilibrate the cartridge with running buffer and

operatate the cartridge according to the CHT
instruction manual or according to your protocol.

4.1 Sample Preparation

Proper pH and ionic strength are necessary for
consistent and reproducible results. Sample can be
exchanged into the starting buffer or diluted to the
starting buffer's concentration. This can be
achieved by diluting the sample to the ionic strength
of the starting buffer, dialyzing against the starting
buffer, or exchanging it into the starting buffer.
Buffer exchange can be accomplished using the
Bio-Scale Mini P6 cartridge, Bio-Spin

®

6 or

Bio-Spin 30 columns, Econo-Pac 10DG desalting
columns, or Bio-Gel

®

P-6DG gel filtration gel. The

choice of product will depend on sample volume.
All samples should be filtered through a 0.45 µm
filter prior to cartridge application.

14

10011433A.qxp 10/30/2007 8:21 AM Page 14