5.2 Wash Alternatives
Perform wash alternatives with any of the following
alternative buffers in place of step 1. Continue with
step 2 listed in Section 5.1.
•
1–2 M KCI or NaCI
•
6 M urea
•
8 M guanidine-HCI
All the wash alternative buffers should contain 5 mM
phosphate at neutral pH.
5.3 Autoclaving
Bio-Scale Mini cartridges are not autoclavable.
5.4 Storage
After the cartridges are washed with deionized
water, Bio-Scale Mini CHT cartridges should be
purged and stored with 0.1 N NaOH. Higher
concentrations of NaOH may be used if desired.
Used CHT, after being regenerated and sanitized,
can be stored in up to 1.0 N NaOH at room
temperature.