Section 5
Care of the Cartridge

5.1 Cleaning

After repeated use, a CHT cartridge may require
thorough cleaning and regeneration to remove
bound contaminants. Most bound contaminants
may be removed by the following protocol:

Step 1. Wash the cartridge with 3–5 column

volume (5 ml/min for the 5 ml) of 500 mM
potassium or sodium phosphate, then
5 column volumes of starting buffer.

Step 2. Equilibrate the cartridge with at least

5 column volumes (5 ml/min for the 5 ml)
of starting buffer. If bound contaminants
persist after following the steps above, use
one of the alternative procedures listed in
wash alternatives.

19

C

CH

HT

T S

Su

uiittaab

biilliittyy

p

pH

H

B

Bu

uffffeerr

1

10

0 ccyycclleess

5

50

0 ccyycclleess

7.0

MES + 5 mM PO4

+

n/a

7.0

Acetate + 5 mM PO4

+

n/a

7.0

Imidazole + 5 mM PO4

+

+

7.0

Glycine + 5 mM PO4

+

n/a

7.0

Arginine + 5 mM PO4

+

n/a

7.0

HEPES + 5 mM PO4

+

n/a

7.0

Tris + 5 mM PO4

+

n/a

7.5

Phosphate (5 mM)

+

n/a

7.5

MES + 5 mM PO4

+

n/a

7.5

Imidazole + 5 mM PO4

+

+

7.5

Acetate + 5 mM PO4

+

n/a

7.5

HEPES + 2 mM PO4

+

n/a

7.5

HEPES + 5 mM PO4

+

n/a

7.5

Tris + 2 mM PO4

+

n/a

7.5

Tris + 5 mM PO4

+

+

8.5

Tris + 5 mM PO4

+

n/a

* All experiments performed in small scale columns. Each cycle used approximately 35 column

volumes of buffer to simulate an equilibration and long gradient, as well as five column volumes

of 1 N NaOH to simulate regeneration.

18

10011433A.qxp 10/30/2007 8:21 AM Page 18