Bio-Rad Bio-Scale™ Mini CHT™ 40 µm Cartridges User Manual

Page 9

Advertising
background image

3.1 BioLogic DuoFlow

Systems

The Bio-Scale Mini cartridge to BioLogic system
fittings kit, (catalog number 732-0113) includes
1/4–28 female to male luer and 1/4–28 female to
female luer to connect one Bio-scale Mini cartridge
to the BioLogic DuoFlow system.

3.2 HPLC Systems

The luer to 10–32 adaptor fittings kit (catalog number
732-0112), provides fittings necessary to connect
the cartridge to nut and ferrule type fittings found
on most HPLC systems. Alternatively, the cartridge
can be connected to HPLC systems via a low-
dead-volume 1/16 inch union with a new piece of
stainless steel tubing attached to the union. Simply
slip a short length of the 0.8 mm ID tubing over
1/16 inch OD stainless-steel tubing to a distance of
1 cm.

11

Section 3
Connecting to Other Liquid
Chromatography Systems

The Bio-Scale Mini cartridges can be connected to
any liquid chromatography system, provided that
the maximum pressure limit (3 bar, 45 psi, or
300 KPa) of the cartridges is not exceeded. It is
recommended that the system pressure limit be set
according to the cartridge pressure limit. Pressures
in excess of 3.4 bar are usually caused by
restrictions in tubing or detector cells downstream
from the cartridge. Bio-Rad offers two fitting kits for
easy connection of a Bio-Scale Mini cartridge to a
BioLogic DuoFlow, HPLC- or FPLC-type system.

10

10011433A.qxp 10/30/2007 8:21 AM Page 10