Integra LifeSciences CryoSolutions Cryosurgical Unit User Manual

Page 5

Advertising
background image

5

performance of the CryoSolutions

®

device but may be

influenced by poor contact or ineffective application times.
Successful outcomes, therefore, are dependent upon
the optimal cold performance delivered by CryoSolutions

®

and application times guided by clinical experience.

• Only use lint-free cloth.

• Sterilize tips at recommended parameters.

See “Cleaning/Sterilizing of Dispensing Tips.”

Adverse Reactions

• Reaction to cryosurgery may differ depending upon the

patient as well as size, type and location of area being
treated. These could range from a barely evident reaction
to an immediate blistering.

Instructions
Cartridge Installation

• Always use protective gloves when installing cartridges.

• Inspect cartridge for o-ring. If o-ring is missing, do

not attach the cartridge. Contact Integra Miltex for

o-ring replacement.

• The pin must be fully inserted into the pin insertion

point before attaching the cartridge.

• Do not use the trigger or any other mechanism to attach

cartridges.

• Contact Integra Miltex for replacement pins.

• Insert the metal pin in the pin insertion point hole on

the control mechanism.

• Hold the control mechanism at the pin position and not at

the on/off trigger. This helps grip the body without

damaging the trigger.

• Care should be taken to align the cartridge and control

mechanism so that gas does not inadvertently escape.

• Screw the cartridge into the control mechanism in a

clockwise direction both quickly and completely to ensure

proper functionality.

• Small amounts of gas escaping from the cartridge while

being attached indicates poor alignment of the cartridge to

the control mechanism.

• Remove the pin.