If the ambient temperature changes during a process and causes the instrument process set temperature
to change beyond the +/- 1

°C specification for more than 2 minutes, a message will appear to inform the

user that the ambient temperature is “high”. A constant beep will sound. The counter will continue to
count. The present temperature will be displayed. Hitting the “Stop” button will allow user to abort the
process. A new screen will be displayed asking user if they are sure they want to abort.

Hyb only:

“Ambient Temp High!” flashes

alternately on “Hyb in Process” line.




Fixed Temp:

“Ambient Temp High!” flashes

alternately on “Fixed Temp” line.




Abort Screen Message:



PGM -- Name
Hyb in Process

Hyb --

°C --:--

Present Temp: --

°C

PGM – Name

Fixed Temp --

°C

Reset Timer 00:00:00
End PGM/Main Menu

ABORTING!!
Are You Sure?
No
Yes – Main Menu