The Hybridizer will automatically cool to hybridization set temperature once denaturation is completed.

“Ambient Temp High!” flashes

alternately on “Present Temp” line.


Hybridization time will count down from the set time once temperature reaches hybridization set point.


Upon program completion Hybridizer will beep five times and the display will show “PROCESS
COMPLETE”
. Hybridization temperature will be maintained until “End PGM/Main Menu” is accepted by
pressing “Enter” button. Before pressing “Enter” button, remove slides for further processing. If “End
PGM/Main Menu”
is not accepted within the first minute of program completion, hybridization time will
start counting the total time at hybridization temperature.


Run a Hyb Only Program

Turn unit on and wait for the Main Menu screen. Cursor highlights “Run a PGM” line. Press “Enter”
button to accept.

With the arrow keys scroll through program numbers 1 to 40 / program names. If no programs have been
saved advance to programming section of this handbook. To accept, press “Enter” button.

Please Wait

Cooling to Hyb 45°C
Present Temp: 58°C

PGM 01 HER2
Hyb in Process
Hyb: 45°C 12:48
Present Temp: 45°C

PGM 01 HER2
PROCESS COMPLETE
Total Hyb Time 21:05
End PGM/Main Menu

Enter PGM no.
or Scroll (arrows)