Create a hybridization only program (hyb only) – Dako Hybridizer User Manual

Page 18

Advertising
background image

18

Dako Hybridizer

|

Handbook

Create a Hybridization Only Program (Hyb Only)

Turn unit on and wait for the Main Menu screen. Use the arrows keys to move the cursor to “Create a
PGM”
and press “Enter” button to accept. With arrow keys move cursor to “Hyb only” line and press
the “Enter” button to accept.

Hybridizer will advance to the next available program number.


Hybridizer allows creating a program name. Cursor highlights the first name character position Use the
arrow keys to move through character set and press “Enter” button to accept the characters. All 10
character positions must be filled. Press “Enter” button to accept blank characters. For numeric
characters use keypad 0-9.

Character set: A-Z; 0-9; period, - and blank (“Enter” button or move arrow).

Cursor advances to “Hyb Temp”. With numeric keypad enter a two-digit temperature value in degrees
Celsius (30-70

°C). The instrument allows a temperature of 30°C or ambient temp + 5°C (whichever is

higher) for the lowest hybridization temperature. For room temperature hybridization (ambient temp +
5

°C) enter the two-digit value 00.


Cursor advances to “Hyb Time”. With numeric keypad enter a two-digit time value in hours (0 – 99)
followed by a two-digit value in minutes (0-59).

Display will now show entered program values. Cursor highlights “Enter to Accept” line.


Press “Enter” button to accept the program values; or press “Backspace” button to return to previous
screen to modify program values; or press “Stop” button to abort.

Note: Accepting the program saves the program by its assigned program number and name.

PGM 02 Name

Enter: Temp Time
Hyb: 30°C 00:00

PGM 02 EBV
Hyb: 55°C 01:30
ENTER to Accept
STOP to Abort