Dako Hybridizer

|

Handbook

15

If the ambient temperature changes after a process is completed, but before the user removes the slides
and causes the instrument process set temperature to change beyond the +/- 1

°C specification for more

than 2 minutes a message will appear to inform the user that the ambient temperature is “high”. A
constant beep will sound. The counter will continue to count. The present temperature will be displayed.
Hitting the “Stop” button will allow user to abort the process. A new screen will be displayed asking user
if they are sure they want to abort.

Hyb only:

“Ambient Temp High!” flashes

alternately on “PROCESS COMPLETE” line.




Abort Screen Message:



If 40 programs have been created or edited the software will blank out the “Create” mode on the main
menu screen. This will only allow users to edit existing programs.





PGM – Name
PROCESS COMPLETE Total
Hyb Time --:-- End
PGM/Main Menu

ABORTING!!
Are You Sure?
No
Yes – Main Menu

Run a PGM
Edit a PGM

Present Temp: --