Dako Hybridizer

|

Handbook

Upon program completion Hybridizer will beep five times and the display will show “Process Complete”.
Hybridization temperature will be maintained until “End PGM/Main Menu” is accepted by pressing
“Enter” button. Before pressing “Enter” button, remove slides for further processing. If “End PGM/Main
Menu”
is not accepted within the first minute of program completion, hybridization time will start counting
the total time at hybridization temperature.

Run a Fixed Temp Program

Turn unit on and wait for the Main Menu screen. Cursor highlights “Run a PGM” line. Press “Enter”
button to accept.

With the arrow keys scroll through the program numbers 1 to 40 / program names. If no programs have
been saved advance to programming section of this handbook. To accept a program number/name,
press the “Enter” button.


Display will confirm PGM number/name and Fixed Temp. Cursor highlights “Run PGM” line.



Press “Enter” button. Instrument will heat to fixed temperature.

PGM 02 EBV
PROCESS COMPLETE
Total Hyb Time 02:15
End PGM/Main Menu

Enter PGM no.
or Scroll (arrows)

PGM 01 HER2

PGM 03 Appl
Fixed: 65°C
Run PGM
Main Menu

Please Wait

Heating to Fxd 65

° C

Present Temp: 47

° C