Troubleshooting

Unit does not turn on, no
power.

Check both cord ends are plugged in.
Check Fuses located on rear panel next to power switch.
Replace with same type and value.

Poor results on slides.

Verify selected protocol against probe manufacturer’s
recommendation.
Ensure Humidity Control Strips are in place and moist.
Ensure cover slip sealant was applied.
Ensure plate is heating.
Ensure lid is properly closed.
Ensure air intake and outlet are not obstructed.
Ensure lid gasket is intact and undamaged

Cannot read display.

Allow unit to reach room temperature before operating.
Call Dako Customer Service

Can’t find “Create a
Program”.

More than 40 programs have been stored. You can only edit
a program.

Keypad not functioning
properly.

Call Dako Customer Service.