3.3.6
Analog Output Menu
If the analog output option is installed, configure all other indicator functions and calibrate the indicator (see
Section 4.0 on page 37) before configuring the analog output. See Section 9.10 on page 59 for analog output
calibration procedures.
Figure 3-15. Analog Output Menu
ALG OUT Menu
Parameter
Choices
Description
SOURCE
GROSS
NET
Specifies the weigh data, gross or net, tracked by the analog output.
OFFSET
0
20
Zero offset in %. Select 0% for 0–10 V / 0-20 mA; select 20% for 2-10 V / 4-20 mA output.
ERRACT
FULLSC
Error action. Specifies how the analog output responds to system error conditions.
Set to full value (10 V or 20 mA)
HOLD
Hold current value
ZEROSC
Set to zero value (0 V or 0 mA) or (2 V or 4 mA)
MIN
000000
0 – 999999
Specifies the minimum weight value tracked by the analog output. Specify a weight value (in
primary units) in the range 0 – 999999.
MAX
000010
0 – 999999
Specifies the maximum weight value tracked by the analog output. Specify a weight value (in
primary units) in the range 0 – 999999.
MINNEG
NO
Yes
YES implies that the value in MIN is a negative value. NO is the default and implies that MIN is
a positive value.
MAXNEG
NO
Yes
YES implies that the value in MAX is a negative value. NO is the default and implies that MAX is
a positive value.
TWZERO
000000
0 – 65535
Tweak zero. Adjust the analog output zero calibration. Use a multimeter to monitor the analog
output value. Press
or
to select the digit to edit.
Press
or
to edit the selected
(flashing) digit. Press
to save the new value.
TWSPAN
000000
0 – 65535
Tweak span. Adjust the analog output span calibration. Use a multimeter to monitor the analog
output value. Press
or
to select the digit to edit.
Press
or
to edit the selected
(flashing) digit. Press
to save the new value.
OUTPUT
CURENT
Defines the output type.
0-20 mA or 4-20 mA Depends on OFFSET
VOLTAG
0-10 V or 2-10 V Depends on OFFSET
Table 3-12. Analog Output Menu Parameters
CONFIG
FORMAT
CALIBR
PROGRM
DIG IN
ALGOUT
PASWRD
TEST
MENU
SOURCE
NO
Yes
NO
YES
OFFSET
ERRACT
MIN
MAX
MINNEG
TWZERO
TWSPAN
MAXNEG
OUTPUT
000000
000000
CURENT
VOLTAG
GROSS
NET
0%
20%
000000
000010
FULLSC
HOLD
ZEROSC