CardioPerfect Workstation

Resting ECG Module - User Manual

DIR 80012333 Ver. E

11 / 54

Precordial electrodes
Connect the precordial electrodes in the following positions:

AHA

IEC

V1

Fourth intercostal space at the right border of the sternum.

C1

V2

Fourth intercostal space at the left border of the sternum.

C2

V3

Midway between locations V2 and V4.

C3

V4

At the mid-clavicular line in the fifth intercostal space.

C4

V5

At the anterior axillary line on the same horizontal level as V4.

C5

V6

At the midaxillary line on the same horizontal level as V4 and V5.

C6

RA

Right arm

R

LA

Left arm

L

RL

Right leg

N

LL

Left leg

F

Reusable Electrodes

Each electrode must be attached securely. The electrode paste, gel, or cream must cover an
area the size of the electrode, but must not extend beyond it, especially on the chest.

LL

V1 V2

V3

V4 V5 V6

RA

LA

RL

N

F

C1 C2

C3

C4 C5 C6

R

L