HRV view (optional)
This tab shows the heart rate variability.
Additional information is available upon
request.

Note:
HRV option not sold in the US.


The view in the workspace can be used in for example a Word document. Right-click in the
Workspace and select copy from the pop-up menu; paste the view in a document.

4.3. Parameters

The lower part of the workspace contains the Parameters pane. The Parameters pane contains
global parameters that are determined from all leads. These parameters are:

Patient age and heart rate

P and PQ duration

QRS duration

QT, QTc and QTd duration

P, QRS and T axis

4.4. QTc

The QT interval in an ECG is a good indicator of the repolarization. However, the QT depends
heavily on the heart rate. To make a useful measurement, the QT is usually corrected.

Welch Allyn CardioPerfect offers three correction methods that can be used to derive the QTc
parameter:

Bazett:

60

60

HR

QT

HR

QT

RR

QT

QTc

B

Fridericia:

3

3

3

60

60

HR

QT

HR

QT

RR

QT

QTc

F

Hodges:

)

60

(

*

75

.

1

HR

QT

QTc

H