Table 3.

Guidance and Manufacturer’s Declaration - Immunity

The Image Capture System is intended for use in the electromagnetic environment specified below. The customer or

user of the Image Capture System should ensure that it is used in such an environment

Immunity Test

IEC 60601

Test Level

Compliance

Level

Electromagnetic Environment - Guidance

Conducted RF
IEC 61000-4-6

Radiated RF
IEC 61000-4-3

3 Vrms
150 kHz to 80MHz

3 V/m
80 MHz to 2.5 GHz

3 Vrms

3 V/m

Portable and mobile communications equipment
should be separated from the EUT by no less than
the distances calculated/listed below:
D=0.35 (Sqrt P)

D=0.35 (Sqrt P)
80 to 800MHz

D=0.7 (Sqrt P)
800MHz to 2.5 GHz
Where P is the max power in watts and D is the
recommended separation distance in meters.
Field strengths from fixed transmitters, as
determined by an electromagnetic site survey,
should be less than 10Vrms or 10V/m.
Interference may occur in the vicinity of
equipment containing a transmitter.

Conducted RFI and Radiated RFI testing was performed on the EUT. Monitoring the output for drops in voltage; none

were detected.

Table 4.

Recommended Separations Distances between portable and mobile RF

Communications equipment and the Welch Allyn Image Capture System

Guidance and Manufacturer’s Declaration - Emissions

The Image Capture System is intended for use in the electromagnetic environment in which radiated

disturbances are controlled. The customer or user of the Image Capture System can help prevent

electromagnetic interference by maintaining a minimum distance between portable and mobile RF

Communications Equipment and the Image Capture System as recommended below, according to the

maximum output power of the communications equipment.

Max Output

Power (Watts)

Separation (m)

150kHz to 80MHz

D=(3.5/V1) (Sprt P)

Separation (m)

80 to 800MHz

D=(3.5/E1) (Sprt P)

Separation (m)

800MHz to 2.5GHz

D=(7/E1) (Sprt P)

0.01

0.035

0.035

0.07

0.1

0.111

0.111

0.221

1

0.35

0.35

0.7

10

1.107

1.107

2.214

100

3.5

3.5

7.0

Advertising