Welch Allyn Suretemp Plus Calibration/Verification Module Notes - Quick Reference Guide User Manual

Page 10

Advertising
background image

Speci

es

Code Definition / Description Origin/Alert

type

Traditional

code map
(ST+ OEM)

Traditio

nal code

map
(ST+

HH)

Details

Host

controller

action

8

37 Under minimum battery

volts.

Power/ Host

controller alert

60

Internal

malfunction.

Contact Welch

Allyn

Technical

Service.

Indicate

module

error

8

38 Battery volts not set.

Module/ Host

controller alert

Internal

malfunction.

Contact Welch

Allyn

Technical

Service.

Indicate

module

error

8

39 Predict algorithm not

set.

Module/ Host

controller alert

Internal

malfunction.

Contact Welch

Allyn

Technical

Service.

Indicate

module

error

8

40 Ambient temperature

not set.

Module/ Host

controller alert

Internal

malfunction.

Contact Welch

Allyn

Technical

Service.

Indicate

module

error

8

41 Non-responsive probe.

Thermistor pulled away

from the tip or heater

broken.

Probe/ Equipment

alert

32/33

E0.5/E0.9 Probe warmer

is not

responding.
Try another

predict cycle.

If the problem

persists,

replace the

probe.

Indicate

probe

error

8

42 Bad probe gain.

Probe/ Equipment

alert

34

E0.9

The probe was

not

characterized/

calibrated.
Calibrate or

replace the

probe.

Indicate

probe

error

8

43 Bad probe response

value.

Probe/ Equipment

alert

34

E0.9

The probe was

not

characterized/

calibrated.

Indicate

probe

error