Code Definition / Description Origin/Alert

type

Traditional

code map
(ST+ OEM)

Traditio

nal code

map
(ST+

HH)

Details

Host

controller

action

F),

respectively.

If conditions

are valid and

the problem

persists,

replace the

probe. If

problem

persists,

replace

module.

8

3

Internal calibration

resistor (RCAL) on the

board is damaged or

contaminated (pulse too

long).

Module/

Equipment alert

47

E4.1

Internal

malfunction.

Contact Welch

Allyn

Technical

Service.

Indicate

module

error

8

4

Internal calibration

resistor (RCAL) on the

board is damaged or

contaminated (pulse too

short).

Module/

Equipment alert

48

E4.1

Internal

malfunction.

Contact Welch

Allyn

Technical

Service.

Indicate

module

error

8

5

Internal circuit

validation resistor (PTB)

on the board is damaged

(value is over).

Module/

Equipment alert

49

E4.0

Internal

malfunction.

Contact Welch

Allyn

Technical

Service.

Indicate

module

error

8

6

Internal circuit

validation resistor (PTB)

on the board is damaged

(value is under).

Module/

Equipment alert

50

E4.0

Internal

malfunction.

Contact Welch

Allyn

Technical

Service.

Indicate

module

error

8

7

A/D measurement timed

out.

Module/

Equipment alert

51

E0.2

Internal

malfunction.

Contact Welch

Allyn

Technical

Service.

Indicate

module

error

8

8

The probe was not

characterized/calibrated.

Probe/ Equipment

alert

52

E0.9

Probe

malfunction;

replace probe.

Indicate

probe

error