Code Definition / Description Origin/Alert
type
Traditional
code map
(ST+ OEM)
Traditio
nal code
map
(ST+
HH)
Details
Host
controller
action
persists,
replace the
probe.
8
31 Excessive heater energy. Probe/ Equipment
alert
22
E0.1
Probe warmer
is not
working. If the
problem
persists,
replace the
probe.
Power
down
8
32 Host controller interface
error.
Module/ Host
controller alert
Internal error
Internal error.
Try another
predict cycle.
If the problem
persists,
replace the
module.
Clear
8
33 Over ambient
temperature.
Ambient/
Equipment alert
24
Device
environment is
too warm.
Move module
to within
ambient
temperature
range.
Clear
8
34 Under ambient
temperature.
Ambient/
Equipment alert
25
Device
environment is
too cold.
Move module
to within
ambient
temperature
range.
Clear
8
35 Invalid SureTemp Plus
algorithm.
Module/ Host
controller alert
Internal Error
Internal error.
Try another
predict cycle.
If the problem
persists,
replace the
module.
Indicate
module
error
8
36 Over maximum battery
volts.
Power/ Host
controller alert
59
Internal
malfunction.
Contact Welch
Allyn
Technical
Service.
Indicate
module
error