Welch Allyn Suretemp Plus Calibration/Verification Module Notes - Quick Reference Guide User Manual

Page 9

Advertising
background image

Speci

es

Code Definition / Description Origin/Alert

type

Traditional

code map
(ST+ OEM)

Traditio

nal code

map
(ST+

HH)

Details

Host

controller

action

persists,

replace the

probe.

8

31 Excessive heater energy. Probe/ Equipment

alert

22

E0.1

Probe warmer

is not

working. If the

problem

persists,

replace the

probe.

Power

down

8

32 Host controller interface

error.

Module/ Host

controller alert

Internal error

Internal error.

Try another

predict cycle.

If the problem

persists,

replace the

module.

Clear

8

33 Over ambient

temperature.

Ambient/

Equipment alert

24

Device

environment is

too warm.

Move module

to within

ambient

temperature

range.

Clear

8

34 Under ambient

temperature.

Ambient/

Equipment alert

25

Device

environment is

too cold.

Move module

to within

ambient

temperature

range.

Clear

8

35 Invalid SureTemp Plus

algorithm.

Module/ Host

controller alert

Internal Error

Internal error.

Try another

predict cycle.

If the problem

persists,

replace the

module.

Indicate

module

error

8

36 Over maximum battery

volts.

Power/ Host

controller alert

59

Internal

malfunction.

Contact Welch

Allyn

Technical

Service.

Indicate

module

error

Advertising