Speci

es

Code Definition / Description Origin/Alert

type

Traditional

code map
(ST+ OEM)

Traditio

nal code

map
(ST+

HH)

Details

Host

controller

action

Technical

Service.

5

24 The module’s power rail

is too low.

Power/

Equipment alert

27

Internal

malfunction.

Contact Welch

Allyn

Technical

Service.

Indicate

module

error

5

25 The module’s power rail

is too low.

Power/

Equipment alert

28

Internal

malfunction.

Contact Welch

Allyn

Technical

Service.

Indicate

module

error

5

26 The reference voltage

circuit was detected to

be under voltage or

unstable.

Module/ Host

controller alert

61

E4.2

Internal

malfunction.

Contact Welch

Allyn

Technical

Service.

Indicate

module

error

8

1

Measurement is below

allowable temperature

values and beyond the

low ambient and/or

patient limits.

Ambient/ User

Alert

45

E0.2

Verify

conditions at

26.67° C (80°

F) and 10°C

(50° F),

respectively.

If conditions

are valid and

the problem

persists,

replace the

probe. If

problem

persists,

replace

module.

Clear

8

2

Measurement above

allowable temperature

values and beyond the

high ambient and/or
patient limits.

Ambient/ User

Alert

46

E0.2

Patient or

environmental

temperature

conditions

may be too

high. Verify

conditions at

43.33° C

(110° F) and

40° C (104°

Clear