Welch Allyn Suretemp Plus Calibration/Verification Module Notes - Quick Reference Guide User Manual
Page 7
Speci
es
Code Definition / Description Origin/Alert
type
Traditional
code map
(ST+ OEM)
Traditio
nal code
map
(ST+
HH)
Details
Host
controller
action
8
16 Cannot read the probe's
EEPROM correctly or
the probe left the factory
without being tested.
Probe/ Equipment
alert
75
E0.7
Verify that
batteries are
fresh. If the
problem
persists,
replace the
probe.
Indicate
probe
error
8
17 Invalid events index.
Module/ Host
controller alert
Event log is
out of range.
Retry.
Indicate
module
error
8
18 Problem reading the
module's EEPROM or
saving to the module's
EEPROM in biotech
mode.
Module/ Host
controller alert
67/68/69
Verify that
batteries are
fresh. Try to
make changes
again. If the
problem
persists,
replace the
module.
Indicate
module
error
8
19 Problem reading the
probe's EEPROM.
Probe/ Equipment
alert
70
E0.6
Verify that
batteries are
fresh. If the
problem
persists,
replace the
probe.
Indicate
probe
error
8
20 TEMP CONFIG
ACQUIRE FAILURE.
Module/
Equipment alert
8
21 TEMP CONFIG
RELEASE FAILURE.
Module/
Equipment alert
8
22 TEMP CONFIG
INVALID PTR
FAILURE.
Module/
Equipment alert
8
23 Internal error. EEPROM
not initialized.
Module/
Equipment alert
74
E4.7
EEPROM not
initialized.
Device will
try to initialize
the module's
EEPROM. If
the problem
persists,
replace the
module.
Indicate
module
error
8
24 Heater indicates on
when turned off.
Module/
Equipment alert
80
E5.0
Heater control
error. If the
Power
down