Welch Allyn Suretemp Plus Calibration/Verification Module Notes - Quick Reference Guide User Manual

Page 7

Advertising
background image

Speci

es

Code Definition / Description Origin/Alert

type

Traditional

code map
(ST+ OEM)

Traditio

nal code

map
(ST+

HH)

Details

Host

controller

action

8

16 Cannot read the probe's

EEPROM correctly or

the probe left the factory

without being tested.

Probe/ Equipment

alert

75

E0.7

Verify that

batteries are

fresh. If the

problem

persists,

replace the

probe.

Indicate

probe

error

8

17 Invalid events index.

Module/ Host

controller alert

Event log is

out of range.

Retry.

Indicate

module

error

8

18 Problem reading the

module's EEPROM or

saving to the module's

EEPROM in biotech

mode.

Module/ Host

controller alert

67/68/69

Verify that

batteries are

fresh. Try to

make changes

again. If the

problem

persists,
replace the

module.

Indicate

module

error

8

19 Problem reading the

probe's EEPROM.

Probe/ Equipment

alert

70

E0.6

Verify that

batteries are

fresh. If the

problem

persists,

replace the

probe.

Indicate

probe

error

8

20 TEMP CONFIG

ACQUIRE FAILURE.

Module/

Equipment alert

8

21 TEMP CONFIG

RELEASE FAILURE.

Module/

Equipment alert

8

22 TEMP CONFIG

INVALID PTR

FAILURE.

Module/

Equipment alert

8

23 Internal error. EEPROM

not initialized.

Module/

Equipment alert

74

E4.7

EEPROM not

initialized.

Device will

try to initialize

the module's

EEPROM. If

the problem

persists,

replace the

module.

Indicate

module

error

8

24 Heater indicates on

when turned off.

Module/

Equipment alert

80

E5.0

Heater control

error. If the

Power

down