SureTemp Plus error codes

To find the SureTemp Plus error code:
1.
Go to the tab of the device you are working on and select the SureTemp Plus module.
2. Select the View log button in the user information section of the page. A .txt document opens with the last 10

SureTemp Plus error codes. The last listed error code is the latest error code.

The module provides these errors in a CErrEStd object. For a unique error code, merge the identified species value

with the error code. If the status indicates a fault condition, the module suggests recovery attempts, which may

include the host initializing a reset.
The errors are grouped into the following origins:
• Probe
• Module
• Ambient
• Power
• Patient
Errors are also classified as the following alert types:
• Host controller
• Equipment
• User

Speci

es

Code Definition / Description Origin/Alert

type

Traditional

code map
(ST+ OEM)

Traditio

nal code

map
(ST+

HH)

Details

Host

controller

action

0

0

No error - used for ACK

messages in response to

commands.

Module/ Host

controller alert

Internal error

None

None

1

5

WACP message CRC

mismatch.

Module/ Host

controller alert

Internal error

None

None

2

1

This message is not

implemented by the

module.

Module/ Host

controller alert

Internal error

None

None

2

2

This message is not

supported by the

module.

Module/ Host

controller alert

Internal error

None

None

2

3

The module has run out

of memory.

Module/ Host

controller alert

Internal error

None

None

2

5

The parameter provided

is invalid for the

specified message.

Module/ Host

controller alert

Internal error

None

None

2

6

The parameter provided

is outside of the

allowable range for the

specified message.

Module/ Host

controller alert

Internal error

None

None

Advertising