Welch Allyn Suretemp Plus Calibration/Verification Module Notes - Quick Reference Guide User Manual

Page 8

Advertising
background image

Speci

es

Code Definition / Description Origin/Alert

type

Traditional

code map
(ST+ OEM)

Traditio

nal code

map
(ST+

HH)

Details

Host

controller

action

problem

persists,

replace the

module.

8

25 Heater indicates off

when turned on.

Module/

Equipment alert

81

E5.0

Heater control

error. If the

problem

persists,

replace the

module.

Indicate

module

error

8

26 HTR_Q on and HTRC is

off, but still has voltage.

Module/

Equipment alert

82

E5.0

Heater control

error. If the

problem

persists,

replace the

module.

Power

down

8

27 HTR_Q is tri-stated with

HTRC enabled and has

heater power.

Module/

Equipment alert

83

E5.0

Heater control

error. If the

problem

persists,

replace the

module.

Power

down

8

28 Turned Q&C on and

heater voltage not high

enough.

Module/

Equipment alert

84

E5.0

Heater control

error. If the

problem

persists,

replace the

module.

Indicate

module

error

8

29 Heater hardware failsafe

should have turned off

but did not.

Module/

Equipment alert

85

E5.2

Heater control

error. If the

problem

persists,

replace the

module.

Power

down

8

30 Probe over 112° F /

43.3° C.

Module/

Equipment alert

21

E0.4

Probe tip over

temperature.

Heater error or

probe

construction

problem.
Assure that

module is

within

ambient temp

limits. If the

problem

Power

down