Chapter 2 Plan the installation
Welch Allyn Wireless Acuity LT Central Monitoring System
Installation plan examples
The following examples show some typical Acuity LT System layouts. These examples are
for illustration only; they are not a guarantee of actual coverage. The network coverage in
each facility depends on the amount and type of obstructions.
Note that although the examples show the switch, PoEs, and a UPS located in a nearby
equipment closet, the equipment closet is optional. This equipment can be located in
either of the following places:
•
Near the Acuity LT Central Station at the nurses station if there is adequate space.
•
As far as 300 feet away from the Acuity LT Central Station. 300 feet is the maximum
length of wiring allowed between the switch and the CPU.
Figure 1. L-shape floor plan
Access point 1
Acuity LT Central
Station
Equipment closet
Access point 2
Allow as much overlap
as possible between AP
areas while still
covering the desired
area in the facility.
AP
AP
Effective network
coverage is limited to
approximately 60 feet,
depending on the
number of walls and
other obstructions.
Confirm coverage in
your facility by testing.
10 feet