Welch Allyn Acuity LT Mobile - Installation Guide User Manual
Page 30

26
Chapter 4 Install the system
Welch Allyn Wireless Acuity LT Central Monitoring System
4.
For one of the following patient monitors, take these steps to set the monitor to the
alternate patient simulation mode:
•
Micropaq monitor:
Press
to
highlight
TOGGLE DEMO MODE, and then press
.
•
Propaq LTR monitor:
a.
Highlight SpO2, and then press
.
b.
Highlight Setup, and then press
.
c.
Highlight Timings, and then scroll to highlight Demo Mode.
d.
Highlight Low, and then press
to return to the main display.
•
Wireless Propaq CS monitor:
a.
Press SETUP, WAVE SELECT, and then INSERVICE.
b.
Confirm that the Waveform Window border changes from red (patient alarm
state) to yellow (equipment alert state).
5.
Click in the waveform area of the Waveform Window to print data from this simulated
patient.
Confirm that within two minutes, the printer prints the data. If it does not, verify that
the printer is online and properly connected.
Print area