Welch Allyn Acuity LT Mobile - Installation Guide User Manual
Page 58

54
Appendix B Check or change a patient monitor network name (ESSID)
Welch Allyn Wireless Acuity LT Central Monitoring System
3.
Press and hold down
and
, and then press
to display the following screen:
4.
Press
to
highlight
YES, and then press
to display the following screen:
The Micropaq monitor displays one of the following pre-set names:
com.protocol
demo.protocol
com1.protocol
com2.protocol
. . ....
com7.protocol
com8.protocol
•
To change the network name, press
or
to highlight the desired network
name, and then press
. The Micropaq monitor turns off, and then turns on and
seeks to connect to the Acuity LT network with the new network name.
If you are not able to change the network name, contact Welch Allyn Technical
Support for assistance (see
•
To exit without changing the network name, make sure EXIT is highlighted and
press .
WARNING When setting the patient monitor network name, make sure that it
matches the network name labeled on the CPU for your wireless Acuity LT
System. If a different network name is set, the monitor could connect to a
different nearby wireless Acuity System. This would result in confusion about
patient monitoring and the location of patient data.