8. Calibration

8.1.

Calibration Group

Calibration

Calibration using a weight that you have

Calibration test

Calibration test using a target mass


Caution

Do not allow vibration or drafts to affect the balance during calibration.
Calibration test does not perform calibration.
To output the data for GLP using the RS-232C interface, set "GLP output (

info

)" of "Data

output (

dout

)". Refer to "10. Function Table".

Calibration test is available only when "GLP output (

info

)" of "Data output (

dout

)" is set .

Caution on using an external weight

The accuracy of an external weight can influence the accuracy of weighing.
Select a mass for calibration and calibration test from the following table.

Model Usable

calibration mass

Adjustable range

HR-300i 100g,

200g*, 300g

-15.0 mg to +15.9 mg

HR-202i 20g,

50g,

100g,

200g*

-15.00 mg to +15.99 mg

The calibration mass in bold type: factory setting

The calibration mass value can be adjusted within the range above.

Display

This indicator means "In process of measuring calibration data".
Do not allow vibration or drafts to affect the balance while the indicator is
displayed.