Details of the function table – A&D Weighing HR-202i User Manual

Page 30

Advertising
background image

28

10.2.

Details of the Function Table

Class

Item and Parameter

Description

0

Fast response, sensitive value

FAST

1

MID.

Cond
Condition

2

Slow response, stable value

SLOW

Can be changed by response

adjustment.

0

Stable range is ±1 digit

1

#1

5t-b
Stability band width

2

Stable range is ±3 digits

The stabilization indicator illuminates

when the display fluctuation is within

the range per second.

0

OFF

1

Normal

2

Strong

trc
Zero tracking

3

Very strong

Keeps zero display by tracking zero

drift.

0

5 times/second

5pd
Display refresh rate 1

10 times/second

Period to refresh the display

0

Point (.)

pnt
Decimal point

1

Comma (,)

Decimal point format

0

OFF

ba5fnc
Environment
Display

p-on
Auto display-ON

1

ON

Turns on the weighing mode display

when the AC adapter is connected.

0

Key mode

Accepts the PRINT key only when

the display is stable.

1

Auto print mode A

(Reference = zero)

2

Auto print mode B

(Reference = last stable value)

Outputs data when the display is

stable and conditions of ap-p, ap-b

and the reference value are met.

prt
Data output mode

3

Stream mode

Outputs data continuously.

0

Plus only

Displayed value>Reference

1

Minus only

Displayed value<Reference

ap-p
Auto print polarity

2

Both

Regardless of displayed value

0

10 digits

1

100 digits

ap-b
Auto print difference

2

1000 digits

Difference between reference value

and displayed value

#2

0

No output

5-id
ID number output

1

Output

Selects whether or not the ID

number is output.

0

No pause

pU5e
Data output pause

1

Pause (1.6 seconds)

Selects the data output interval.

0

Not used

at-f
Auto feed

1

Used

Selects whether or not automatic

feed is performed.

0

No output

1

AD-8121 format

info
GLP output

2

General data format

Selects GLP output method.

For how to set time and date to be

added, refer to peripheral manual.

0

Not used

dout
Data output

ar-d
Zero after output

1

Used

Adjusts zero automatically after data

is output

: Factory settings.

#1 The unit of minimum display is digit. Example: If 1 mg display is selected using the

RANGE

key for the HR-300i, 1 mg is one digit.

#2 Usable minimum display of the balance is one digit. Example: In gram display, one digit is

0.1 mg for the HR-300i , and 0.01 mg for the HR-202i.