11. ID Number and GLP Report

The ID number is used to identify the balance when Good Laboratory Practice (GLP) is used.
The ID number is maintained in non-volatile memory even if the AC adapter is removed.
The GLP output format is selected at "GLP output (

info

)" of the function table and can be

output to a personal computer or printer using the RS-232C serial interface.

The GLP output format includes the balance manufacturer, model, serial number, ID

number, date column, time column and a space for signature for the weighing data, the
weight used and results for calibration or calibration test data. When connecting the
AD-8121B printer, the date and time can be printed using the clock function of the
AD-8121B.

The balance can output the following reports for GLP.

"Calibration report" of the calibration, using an external weight.

"Calibration test report" of the calibration test, using an external weight.

"Title block" and "End block" for the weighing data.


11.1.

Setting the ID Number

1 Press and hold the

RANGE

key until

ba5fnc

of the function table is displayed, then

release the key.


2 Press

the

RANGE

key several times to display

id

.


3 Press

the

PRINT

key. Set the ID number using the following keys.

RE-ZERO

key..... The key to set the character of the digit selected.

Refer to the display character set shown below.

RANGE

key........ The key to select the digit to change the value.

PRINT

key ......... The key to store the new ID number and display

ba5fnc

.

CAL

key............. The key to cancel the new ID number and display

ba5fnc

.


4 With

ba5fnc

displayed, press the

CAL

key to return to the weighing mode.