14. Maintenance

14.1.

Treatment of the Balance

Clean the balance with a lint free cloth that is moistened with warm water and a mild

detergent.

Do not use organic solvents to clean the balance.
Do not disassemble the balance. Contact the local A&D dealer if the balance needs

service or repair.

Use the original packing material for transportation.
Consider section "3. Precautions" for using the balance.

14.2.

Error Codes

Display Error

code

Description

EC,E11

Stability error
The balance can not stabilize due to an environmental
problem. Check around the pan. Prevent vibration, drafts,
temperature changes, static electricity and magnetic fields,
from influencing the balance.
To return to the weighing mode, press the

CAL

key.

EC,E20

Calibration weight error
The calibration weight is too heavy. Confirm the calibration
weight value. Press the

CAL

key to return to the weighing

mode.

EC,E21

Calibration weight error
The calibration weight is too light. Confirm the calibration
weight value. Press the

CAL

key to return to the weighing

mode.

Overload error
A sample beyond the balance weighing capacity has been
placed on the pan. Remove the sample from the pan.

Weighing pan Error
The weight value is too light.
Confirm that the weighing pan is properly installed and
calibrate the balance.

Sample mass error
The balance can not store the sample for the counting mode
or for the percent mode because it is too light.
Use a larger sample.