Display Error

code

Description

Unit mass error
The sample unit mass for the counting mode is too light.
Storing and using it for counting will cause a counting error.
Add samples to reach the specified number and press the

PRINT

key. Pressing the

PRINT

key without adding samples

will shift the balance to the counting mode. But, for accurate
counting, be sure to add samples.

EC,E00

Communications error
A protocol error occurred in communications.
Confirm the format, baud rate and parity.

EC,E01

Undefined command error
An undefined command was received.
Confirm the command.

EC,E02

Not ready
A received command can not be processed.
Example:

The balance received a "Q" command, but not in the

weighing mode.

The balance received a "Q" command while processing a

RE-ZERO command. Adjust the delay time to transmit a
command.

EC,E03

Timeout error
If the timeout parameter is set to "

t-Up1

", the balance did not

receive the next character of a command within the time limit
of one second. Confirm the communication.

EC,E04

Excess characters error
The balance received excessive characters in a command.
Confirm the command.

EC,E06

Format error
A command includes incorrect data.
Example:

The data is numerically incorrect.

Confirm the command.

EC,E07

Parameter setting error
The received data exceeds the range that the balance can
accept. Confirm the parameter range of the command.

Other error code

If an error described above can not be released or other
errors are displayed, contact the local A&D dealer.