11.2.

GLP Report

Set the following parameters to output the report.

Refer to "13.2. Connection to peripheral equipment", for connection to an AD-8121B.
To print the report, set the "GLP output (

info

)" parameter to "

1

" and use MODE 3 of the

AD-8121B. If MODE1 is used, select temporary dump print mode by pressing the

STAT.

key of the AD-8121B.

To output the report to a personal computer using the RS-232C interface, set the "GLP

output (

info

)" parameter to "

2

".

If the time and date are not correct, adjust the settings of the AD-8121B.

Calibration report using an external weight

Key operation
1 Press and hold the

CAL

key to display

Cal out

and release the key.

2

Cal 0

is displayed.

3 When updating the calibration mass value, press the

RANGE

key and proceed to step 4.

When using the preset calibration mass value, proceed to step 5.

4 Specify calibration mass value using the following keys.

RANGE

key ........ The key to select the blinking figure

RE-ZERO

(+)key .. The key to increase the value of the blinking figure.

MODE

(-)key........ The key to decrease the value of the blinking figure.

PRINT

key ......... The key to store the new value.

5 When pressing the

RE-ZERO

key, the zero point is measured and the weight value is

displayed for a few seconds.

6 Place the displayed mass on the pan. Press the

PRINT

key to measure the mass and the

weight value is displayed for a few seconds.

7 If GLP output is used,

glp

is displayed and the calibration report is output.

8 The balance returns to the weighing mode after this calibration.

Setting of "info 1"

Setting of "info 2"

AD-8121 printer format

General format


Space, ASCII 20h

<TERM>

Terminator, CR , LF or CR

CR

Carriage return, ASCII 0Dh

LF

Line feed, ASCII 0Ah

Manufacturer

Model

Serial number

ID number

Date

Time

Calibration type

Calibration weight

Signature

-----------

A

-

&

-

D

MODEL

----

HR-300i

S/N

-----

01234567

ID

------

LAB-0123

DATE

--

2004/09/30

TIME

----

12:34:56

CALIBRATED(EXT.)
CAL.WEIGHT

----

+200.0000

--

g

SIGNATURE


-

-

-

-

-

-

-

-

-

-

-

-

-

-

-

The space

to fill date in.

The space