background image

29

Class

Item and Parameter

Description

0

600

bps

1

1200

bps

2

2400

bps

3

4800

bps

4

9600

bps

bp5
Baud rate

5

19200 bps

0

7 bits, even

1

7 bits, odd

btpr
Data bit, parity bit

2

8 bits, none

0

CR LF

Crlf
Terminator

1

CR

CR: ASCII code 0Dh

LF: ASCII code 0Ah

0

A&D standard format

1

DP format

2

KF format

3

MT format

4

NU format

type
Data format

5

CSV format

Refer to "10.5. Description of Item

"Data Format".

0

No limit

t-Up
Timeout

1

1 second

Selects the wait time to receive a

command.

0

No output

erCd
AK, Error code

1

Output

AK: ASCII code 06h

0

Not used

5if
Serial
interface

Ct5
CTS, RTS control

1

Used

Controls CTS and RTS.

Unit
Unit

Refer to "5. Weighing Units".

id
ID number setting

Refer to "11. ID Number And GLP Report".

: Factory settings. Digit is a unit of minimum weighing value.

Caution

The balance may not transmit the data completely at the specified refresh rate, depending