Flintec PC6D User Manual

Page 29

Advertising
background image

Manual PC6D CANopen Page 29

Zero tracking will be performed only on results less than +/-0.5 d at a rate of 0.4 d/sec, where
d = display step size (see DS command). The zero can only be tracked to +/- 2% of maximum
(see CM command).

Factory default: ZT=0

FD

Factory default settings

[ 2006sub02 ]

This command puts the PC6D back to a known state. The data will be written to the
EEPROM and the TAC will be incremented by 1.

Note: All calibration and setup information will be lost by issuing this command!

Only the factory temperature correction table and serial number are preserved.

Master (PC / PLC) sends

Device responds

Result

CE

E+00017 (example)

Request: TAC-counter CE 17

CE_17 OK

Calibration sequence active

FD OK

Factory default setting


CS

Save the calibration values

[ 2004sub02 ]

This command results in the calibration values being saved to EEPROM, and causes the
TAC to be incremented by 1.

Master (PC / PLC) sends

Device responds

Result

CE

E+00017 (example)

Request: TAC-counter CE 17

CE_17 OK

Calibration sequence active

CS OK

Calibration values saved

The CS command saves all of the calibration group values, as set by CZ, CG, CM, DS, DP
and ZT. The command returns ERR and has no updating action unless it is preceded by the
CE_XXXXX.

Advertising