Manual PC6D CANopen Page 45

5.11 Save calibration, setup & setpoint parameters Commands – CS, WP, SS

The setup and calibration parameters can be divided into 3 groups:

Calibration parameter: CZ, CG, DS, DP & ZT are saved by the CS command.

Setup parameters (other than setpoint): FL, FM, NR, NT, BR, AD, etc. are saved by the

WP command.

Setpoint parameters: Sn, Hn and An are saved by the SS command.

CS

Save the calibration parameters

[ 2004sub02 ]

Note: Calibration parameters can only be saved if the TAC code is known and precedes the
CS command. See the CE and CS commands on

page 27 / 29.

Both the setup parameters and the setpoint parameters are stored in EEPROM using the WP
and SS commands respectively.

WP

Save the setup parameters

[ 2004sub03 ]

With this command the settings of the “Filter” (FL, FM) , the “No-Motion” (NR, NT) and the
communication (AD, BR) will be saved in the EEPROM.

Master (PC / PLC) sends

Device responds

Result

WP OK

Parameter saved

WP ERR

Error


SS

Save the “setpoint” set-up parameters

[ 2004sub05 ]

With this command the settings of the setpoints (Sn), the “setpoint hysteresis” (Hn) and the
“setpoint action” (An) will be saved in the EEPROM.

Master (PC / PLC) sends

Device responds

Result

SS OK

Parameter saved

SS ERR

Error