Manual PC6D CANopen Page 39

5.9 Trigger Commands – SD, MT, GA, TE, TR, TL

Note: All setups should be stored with the WP command before power off.

SD Start

Delay

[ 2500sub0A ]


Set the delay between falling/rising edge of trigger pulse and start of measurement.
Permitted values are 0 … 65535 ms.

Master (PC / PLC) sends

Device responds

Result

SD S+00100

Request: SD=100 ms

SD_200 OK

Setting: SD=200 ms

Factory default: 0 [= 0 ms]


MT

Measuring Time

[ 2500sub09 ]


Set the time over which the average value will be built.
Permitted values are 0 …65535 ms.

Master (PC / PLC) sends

Device responds

Result

MT M+00100

Request: MT=100 ms

MT_500 OK

Setting: MT=500 ms

Note: MT = 0 means disabled trigger and average function.

Factory default: 0


GA Get

Average

[ 2900sub06 ]


Issuing the GA command the PC6D returns the latest average weight value by using the MT
setup. On PC6D it is not necessary to ask for this. The average result is sent in a TPDO2
when ready, or re-triggered.

Master (PC / PLC) sends

Device responds

Result

GA A+01.100

Answer: GA=1.100 g


Note: During the time between the trigger condition being accepted and the average
value being updated, the GA command will return the value 99999 when it has been
triggered or 88888 when it has been retriggered or 99996 when the system tried to
change Tare or Zero before end of measurement.

[ Remark: When started this function, the latest average weight is availabe in TPDO2. ]