5.6 Output

Commands

– GG, GN, GT, GS

The following commands “Get” the Gross, Net, Tare and ADC (Sample) values from the
PC6D. Available on the CAN profile index 2000 and 2001 in floating point and integer
respectively.


GG

Get Gross value

[ 2000/2001sub01 ]

Normally sent in TPDO1, if selected by index 1A00sub01.

Master (PC / PLC) sends

Device responds

Result

GG G+01.100

Gros weight 1.100 d


GN

Get Net value

[ 2000/2001sub02 ]

Normally sent in TPDO1, if selected by index 1A00sub01.

Master (PC / PLC) sends

Device responds

Result

GN G+01.000

Net weight 1.000 d


GT

Get Tare value

[ 2000/2001sub03 ]

Master (PC / PLC) sends

Device responds

Result

GT T+00.100

Tare weight 100 d


GS

Get ADC Sample value

[ 2900sub07 ]

This command gets the actual Analogue to Digital Converter (ADC) value. This can be useful
during development or when calibrating to see how much of the ADC range is being used.

Master (PC / PLC) sends

Device responds

Result

GS S+125785

AD-value = 125.785 d

For service applications, it is helpful to note the GS values for the “no-load” or “zero” output
and when the “calibration load” is applied.