background image

FlintWeigh Technical Manual, Rev. 3.03 November 2011

Page 10 of 32

4.5. Operation of Standard Load Cells

The scale interface can be connected with up to two LDU amplifiers. Depending on the application – legal for
trade scale or industrial application – it can be selected among various weighing modules.

Weighing module type

LDU 68.1

LDU 69.1

LDU 78.1

LDU 68.2

Accuracy class

Not approved

Certificate no.

DK0199-R76-

10.04 Rev.1

DK0199-R76-

11.08 Rev.1

DK0199-R76-

10.08 Rev. 1

Intervals "n" acc. Kompmodeng:

n

ind

10.000

10.000

10.000

Internal resolution

± 130 000

> ± 1 000 000

± 260 000

± 130 000

Approved single range scale

yes

yes

yes

no

Approved multi range scale

no

yes

yes

no

Approved multi interval scale

no

yes

yes

no

Load cell excitation

U

EXC

5 V DC

5 V AC

5 V DC

5 V DC

Minimum input voltage

U

MIN

0 V

0 V

0 V

0 V

Approved min. input per vsi

U

min

1.0

µ

V

0.1

µ

V

0.3

µ

V

Input sensitivity

0.05

µ

V/d

0.02

µ

V/d

0.05

µ

V/d

0.1

µ

V/d

Min. load cell impedance

R

Lmin

87.5

87

87.5

87.5

Max. load cell impedance

R

Lmax

1150

1200

1150

1150

Min. operation temperature

T

min

-15 °C

-15 °C

-15 °C

-15 °C

Max. operation temperature

T

max

+55 °C

+55 °C

+55°C

+55 °C

Fractional error limit

P

ind

0.5

0.5

0.5

Load cell connection

4-/6- wire technique

4/6

4/6

4/6

4/6


Tab. 4.3: Technical characteristics of LDU amplifiers

The connection of the standard load cells has to be done as described in the manual of the KAL-4 digital
junction box (see Flintec-CD, technical documents, manuals).

4.6. Sealing Details with Sticker

The legal for trade operation requires a sealing of the junction boxes. The following figure shows a possible
positioning of the sticker for KAL-4

Fig. 4.6: Sealing details with sticker