FlintWeigh Technical Manual, Rev. 3.03 November 2011

Page 18 of 32

7.3. Scale with Standard Load Cells

Important: The sensor type already has been set to “LDU” (see Fig. 8.2; Register “Communication“).
If you have to configure a multi scale system, then select the wished scale no. ("1" or "2") within the weight
display (Fig. 7.1).

7.3.1.

Define the Scale Parameters

Fig. 8.7 Register ”Calibration“
For sensor type “LDU“ and
single-interval scale















Selected scale


Sensor address:

Defines the address for the LDU (see table below for the factory settings) and enables
the LDU (by activating the check box besides the input field)

Note: The address in parentheses will be used for the second LDU, if two LDUs of the
same type will be used.

LDU-Typ

Sensor address

Comment

68.1

1 (2)

Legal for trade

69.1

3 (4)

Legal for trade

78.1

5 (6)

Legal for trade (see chapter 4.4)

68.2

7 (8)

Not approved

Scale type:

Setting

Comment

Single-interval

The scale has exactly one weighing range with a constant scale
interval

Multi-range

The scale has 2 or 3 weighing ranges with different scale intervals. If
the current gros value of the measured weight falls within the next
weighing range and larger scale interval (changeover threshold 1 and
2), then the larger scale interval remains active until the scale will be
unloaded and will return to the zero point.

Multi-interval

The scale has 2 or 3 weighing ranges with different scale intervals.
The current gros value of the measured weight determines the active
weighing range and the active scale interval at any time.

Scale interval 1/2/3: Step size of the display resolution for the scale respective the weighing range

Decimals:

Number of digits after the decimal point

Unit:

Weight unit for the displayed weight: g, kg, t or lb

No-motion range /
period:

Condition for a stable scale (no. of scale intervals, no. of measured values);
Default: 2d and 10 values in legal for trade applications; in industrial applications these
criteria may be modified

Max. capacity:

Maximum load of the scale (as a multiple of the scale interval e1)