Flintweigh ii configuration, Basic setup during the commissioning – Flintec II PC-BASED User Manual

Page 15

Advertising
background image

FlintWeigh Technical Manual, Rev. 3.03 November 2011

Page 15 of 32

7.

FlintWeigh II Configuration

7.1. Basic Setup during the Commissioning

When you commission FlintWeigh II for the first time, following basic setup has to be done:

Sensor interface and Sensor type (

standard load cells OR digital load cells)

Scale system

Data memory

Interface to remote display and remote display type

Gravitational factor, if applicable

These basic settings will be described in this chapter.

Fig. 8.1 Weight display before commissioning

Start the main programme ”NW11F.EXE“

A click on this button will open the service functions
(window FlintWeigh II Settings).

Fig.8.2 Register „Communication“


Sensor interface:

Select the serial interface, which has been assigned to the weighing system
controller (see chapter 5.3) and set up the baud rate to „9600“

Sensor type:

Defines, if standar or digital load cells will be connected
(LDU = standard load cells; RC3D,= digital load cells)

Scale system:

Defines, if there will be “1 scale“, “2 scales“ (independend from each other) or a
“Dual scale”

(Dual scale = sum of scale no.1 and scale no.2)

Data memory:

Location for the alibi memory (see chapter 8.1.1)

Remote display:

Define interface and type as required

Gravitational factor:

Is only used, if the calibration location and the installation location are not
identical. Otherwise this parameter remains untouched.
Default setting: 9.8075656