Notice – VACUUBRAND MZ 1C User Manual

Page 19

Advertising
background image

page 19 of 102

possible an exchange of gas, albeit extremely slight,

between the environment and the vacuum system.

Adopt suitable measures to prevent contamination of

the pumped substances or the environment.

• Ensure that no parts of your clothing, hair or fingers

can be caught or drawn in at the inlet of the pump.

Never insert fingers or drop any other object into the

inlet or outlet.

• Pay attention to the safety symbol ”hot surfaces” on

the equipment. Hot parts may cause burns if touched.

Adopt suitable measures to prevent any danger aris-

ing from hot surfaces or electric sparks. Ensure that

hot surfaces of the pump do not cause burns. Provide

a suitable contact guard if necessary.

• Pumping at high inlet pressure may lead to overpres-

sure at the gas ballast valve. Pumped gases or con-

densate might be expelled if the valve is open. If an in-

ert gas supply is connected to the gas ballast, ensure

that its inlet pipeline is not contaminated.

Prevent the backpressure of gases and the backflow of

condensates at the outlet.

Provide appropriate protective measures to allow for the

possibility of failure and

malfunction. The protective mea-

sures must also allow for the requirements of the respec-

tive application.

In case of overload, the motor is shut down by a

self-hold

thermal cutout in the winding.

Note: Only manual reset is possible. Switch off the pump

and disconnect from the power source. Identify and elimi-

nate the cause of failure. Wait approx. five minutes before

restarting the pump.

Note: In case of supply voltage below 100V, the lock

of the cutout might be impaired and the pump may

restart on its own after sufficient cooling down. Take

NOTICE