• If the pump is installed at an altitude of more

than 6500 ft (2000 m) above mean sea lev-

el, check compatibility with applicable safety

requirements, and adopt suitable measures.

There is a risk of the motor overheating due

to insufficient cooling.

max. 6500 ft

(2000 m)

• Check compatibility with the

maximally per-

mitted pressure at inlet and outlet and the

maximum pressure difference between

inlet and outlet ports.

max.

16.0 psi

(1.1 bar)

absolute

Do not start the pump if the

pressure difference between

inlet and outlet ports exceeds max. 16.0 psi (1.1 bar).

Attempts to start the pump at higher pressure difference

may cause stalling and damage of the motor.

NOTICE

If pumping condensable vapors (water vapor,

solvents, etc.), let the pump (MZ 1C) run with

gas ballast to help purge any condensation in

the pump.

gas ballast

NOTICE

Prevent internal condensation, transfer of liq-

uids or dust. The diaphragms and valves will

be damaged, if liquids are pumped in signifi-

cant amounts.

Check the pump regularly for external soiling

and deposits. Clean the pump if necessary

to avoid an increase of the pump’s operating

temperature.

NOTICE