Connecting the outlet

(*)

NOTICE

outlet

Outlet via hose nozzle for tubing I.D. 3/8” (hose

nozzle DN 10 mm) (ME 1C, MZ 1C) or via si-

lencer (ME 1).

Attention: Use the silencer only in case of low

gas throughput. Check the silencer regularly for

permeability! In case, install a hose nozzle instead

(*, see “Accessories / Spare parts”, pg. 39).

Connecting an exhaust line at the pump outlet (ME 1):

Unscrew the silencer and install a hose nozzle DN 10/6

mm (thread G1/8”).

If dangerous or polluting fluids could be released at

the outlet, install an appropriate system to catch and

dispose of those fluids.

+

Connect a gas-tight exhaust line at the pump

outlet if necessary. Always vent exhaust

gases appropriately (e.g., into a fume hood).

+

Never block the gas outlet. The exhaust

line must always be free of obstructions (no

back pressure) to ensure an unimpeded

discharge of gas. The cross-section of the

outlet tubing must be at least the size of the

pump’s exhaust connection.

• Reduce the transmission of vibration. Prevent mechan-

ical load due to rigid pipelines. Insert elastic hoses or

flexible elements as couplings between the pump and

rigid pipes.

Secure hose connections at the pump ap-

propriately, e.g., with hose clamps, to protect

against accidental detachment.

To reduce pump noise emanating from the pump

exhaust port, connect an exhaust hose or use

a silencer (see “Accessories / Spare parts”, pg.

39).