Avoid overheating (e.g., due to hot process

gases). Maximum permissible temperature

range: see “Gas inlet temperatures”, pg. 24.

In case of overload, the motor is shut down by

a

self-hold thermal cutout in the winding.

Note: Only manual reset is possible. Switch

off the pump and disconnect from the power

source. Identify and eliminate the cause of fail-

ure. Allow the pump to cool down sufficiently

before restart.

°C / °F

I ➨ O

3.

2.

1.

NOTICE

!

Note: In case of supply voltage below 100V, the lock

of the breaker may not latch and the pump might re-

start on its own after sufficient cooling. Take appropri-

ate precautions, if an automatic restart of the pump

may lead to a dangerous situation.

Make sure the fan’s air supply is adequate.

Check fan regularly for dust/dirt. Clean fan

guard grill if necessary to avoid a reduction of

ventilation.

NOTICE

A warm up period (approximately 15 min.) is

required to ensure that the rated ultimate vac-

uum, the pumping speed, and the full vapor

pumping rate (MZ 1C) are attained.