5

T20

4x

+

2.2 ft.lbf

(3 Nm)

Important: Check operability and safety of the pump

after repair and after replacing the device fuses.

Check the electrical safety (protective conductor resis-

tance, insulating resistance, high voltage test) according

to IEC 61010 and national regulations.