VACUUBRAND MZ 1C User Manual

Page 23

Advertising
background image

page 23 of 102

• The equipment is designated for an ambient and gas inlet temperature

during operation of +10 to +40°C. Never exceed these ambient and

gas inlet temperatures. If pumping / measuring gases which are not

potentially explosive, extended gas inlet temperatures are permissible.

See instructions for use, section “Gas inlet temperatures” or “Techni-

cal data”.

After any intervention at the equipment (e.g., repair / maintenance) the ul-

timate vacuum of the pump has to be checked. Only if the pump achieves

its specified ultimate vacuum is the pump’s leak rate low enough to en-

sure that no explosive atmospheres will occur in the interior of the equip-

ment.

After any intervention at the vacuum sensor, the leak rate of the equip-

ment has to be checked.

Attention: This manual is not available in all

languages of the EU. The user must not op-

erate the device if he does not understand

this manual. In this case a technically cor-

rect translation of the complete manual has

to be available. The manual must be com-

pletely read and understood before opera-

tion of the device. All specified measures

must be applied, or else must be replaced

by equivalent measures at the user’s own

risk.

Advertising
This manual is related to the following products: