Important notes regarding the use of gas ballast

Gas ballast is a continuous purge to keep the pump’s in-

terior as clean as possible and to reduce the possibility of

condensation inside the pump.

Air and pumped media might react inside the pump or

at the outlet of the pump and form hazardous or explo-

sive mixtures, when you use air rather than inert gas

for the gas ballast. This constitutes a risk of significant

damage to equipment and/or facilities, a risk of per-

sonal injury or even loss of life.

+

Make sure that air/gas intake through the gas ballast

valve can never lead to hazardous, explosive or other-

wise dangerous mixtures. If in doubt, use inert gas.

To reduce condensation in the pump, do not pump vapor

before the pump has reached its operating temperature.

Open the gas ballast valve when pumping condensable

vapors. Turn gas ballast cap to open valve.

NOTICE

gas ballast (open)

For

condensable vapors (water vapor, sol-

vents, etc.):

- The gas ballast valve is open if the arrow on

the gas ballast cap is pointing towards the

ON/OFF switch.

- With gas ballast valve open, the ultimate

vacuum will be reduced.

- Use inert gas for gas ballast to avoid the formation of

explosive mixtures.

Attention: maximum supply pres-

sure of inert gas: 17.5 psi (1.2 bar) absolute.

- Close the gas ballast valve by turning the cap 180°.

In case of low boiling solvents (when the formation of con-

densate is unlikely), the use of gas ballast might be un-

necessary.