Switch off the pump.
➨
Disconnect the electrical power cord before remov-
ing the covers. After disconnecting from power,
wait
five seconds to allow the capacitors to discharge.
+
The replacing of the fuses has to be carried out by a
trained electrician.
+
After replacing the fuses, the pump must be checked
for electric safety (see below)! Identify and eliminate the
cause of failure before switching on the pump again.
Replacing the fuses
1
2
T20
4x
+
Attention:
The electrical
system (wires
and capacitors) is
integrated inside
the cover!